Flexcon dermaflex
NWP 240 White H-778 90 PFW

documents

Technical Data sheet

Features and Benefits
    • Soft non-woven polyester fabric is porous to allow maximum breathability
    • H-778 high-performance adhesive with ultra-high strength for extended wear, such as attachment of long-term wearable devices
    • H-778 adhesive bonds to smooth and textured TPO’s and other low-energy surfaces commonly found in medical devices
    • Backed with a 90-lb. moisture stable polycoated layflat release liner suitable for roll form, roll to sheet, and sheet form converting

technical data

Values
Thickness (Mils[microns])
Film
18.0 (457) +/- 15 %
Adhesive
1.9-2.1 (48-53) +/- 0.1 (3)
Liner
6.9 (175) +/- 10 %

Adhesion properties

Ultimate Peel From
Average Oz/In
(N/m)
Acrylic
77
(748)
Glass
64
(704)
Polyethylene
65
(715)
Polypropylene
65
(715)
Stainless Steel
68
(748)

Electrical properties

Property
Value
Test Method
Expected Shear
Room Temp (hours): Less than 1.0
ASTM D 3654 Method A, a. 1 hr. dwell, b. 1 sq. in. surface, c. 4 lb. load
Tack (gm/sq cm)
440
ASTM D 2979
Storage Stability
Two years stored at 70°F (21°C) and 50% relative humidity

Product ID # FLX068292

All of the descriptive information, the typical performance data, and recommendations for the use of FLEXcon products shall be used only as a guide and do not reflect the specification or specification range for any particular property of the product. Furnishing such information is merely an attempt to assist you after you have indicated your contemplated use and shall in no event constitute a warranty of any kind by FLEXcon. All purchasers of FLEXcon products shall be responsible for independently determining the suitability of the material for the purpose for which it is purchased. No distributor, salesman, or representative of FLEXcon is authorized to give any warranty, guarantee, or make any representation in addition or contrary to the above.